917X Firmware

9170 Firmware (.zip)

Additional Downloads:

9171 Firmware (.zip)

9172 Firmware (.zip)

9173 Firmware (.zip)

Title:
917X Firmware
Versions:
9170: Ver. 1.20
9171: Ver. 1.20
9172: Ver. 1.20
9173: Ver. 1.20
Use with:
All 9170, 9171, 9172 and 9173
Date Modified:
1/1/2012

Download Requirements:          

Locate a computer with an available RS-232 COM port, with Windows XP or newer operating system.
Locate an RS-232 Null modem cable and connect one end of the cable to the instrument, the other end to the RS-232 serial port on the computer. 

Note: If your computer only has USB serial interface, an adapter from USB to RS-232 will be required. Recommended adapter: Keyspan USA-19HS.

Instructions:
Select the appropriate download file above; select by Model.

Copy the Code Transfer Software to a convenient directory on the computer. There are three files included in the folder (these include one installation instruction, one .bin file and one .exe to install the program in your unit.  The number in the name of the .bin files must match the model of the instrument to be programmed.

Unzip the files, and read the download instructions.
Run the Code917x.exe (FW download utility).  The software will allow you to select the available COM port.

The New firmware will be transferred to the instrument through the RS-232 cable.  The status of the transfer will be shown on the instrument and on the computer screen.  When complete the computer screen will show “Code Transfer Successful” and the instrument will restart.


Having successfully updated the embedded firmware check to make sure all calibration parameters are correct before making measurements with the instrument.  Calibration parameters can be found on the calibration report for the instrument.
If you have problems with the download or questions about this release, please call customer service at 1-877-355-3225 or email: temperaturesupport@flukecal.com

Converse com oAssistente da Fluke
Limpar conversa
[object Object]