PM600 Pressure Measurement Modules

  • PM600
  • PM600
  • PM600
  • PM600

Informações gerais sobre o produto: PM600 Pressure Measurement Modules

Key features:

  • Fluke Calibration Quartz Reference Pressure Transducer (Q-RPT) technology provides precise measurement with long term stability
  •  0.01% reading measurement uncertainty from 30% to 100% of the modules’ span allows for extremely wide workload coverage
  • Onboard barometer included with absolute mode pressure modules, allowing them to be used in both absolute and gauge mode

Especificações: PM600 Pressure Measurement Modules

Performance specifications

The performance specifications describe the complete instrumental uncertainty of the Product. The specifications include all relevant error components (linearity, hysteresis, repeatability, resolution, reference standard measurement uncertainty, 1-year drift, and temperature effects). The specifications are provided at a level of confidence of 95 %, k=2, normally distributed. Precision uncertainty includes linearity, hysteresis, repeatability, resolution, and temperature effects.


PM600 Modules measurement specifications
Specifications are valid from 15 °C to 35 °C. 
 

ModelAbsolute Mode Range
(SI Units)
Absolute Mode Range
(Imperial Units)
Gauge Mode Range 3
(SI Units)
Gauge Mode Range
(Imperial Units)
1-Year
Instrumental Uncertainty
(% of reading or % FS, whichever is greater)
Precision Uncertainty
(% of reading or % FS, whichever is greater)
BRM600-BA100K470 kPa to 110 kPa10 psi to 16 psi--0.01 % of reading0.008 or 0.0024
PM600-BG15K5---15 kPa to 15 kPa-60 inH2O to 60 inH2O0.01 or 0.0030.008 or 0.0024
PM600-G100K6--0 kPa to 100 kPa0 psi to 15 psi0.01 or 0.0030.008 or 0.0024
PM600-G200K6--0 kPa to 200 kPa0 psi to 30 psi0.01 or 0.0030.008 or 0.0024
PM600-A100K56 kPa to 100 kPa0.9 psi to 15 psi-94 kPa to 0 kPa-13.8 psi to 0 psi0.01 or 0.0031,40.008 or 0.0024
PM600-A200K610 kPa to 200 kPa1.5 psi to 30 psi-90 kPa to 100 kPa-13.2 psi to 15 psi0.01 or 0.0031,40.008 or 0.0024
PM600-A350K610 kPa to 350 kPa1.5 psi to 50 psi-90 kPa to 250 kPa-13.2 psi to 35 psi0.01 or 0.00310.008 or 0.0024
PM600-A700K718 kPa to 700 kPa2.6 psi to 100 psi-82 kPa to 700 kPa-12.1 psi to 100 psi0.01 or 0.00310.008 or 0.0024
PM600-A1.4M70.035 MPa to 1.4 MPa5 psi to 200 psi-0.065 MPa to 1.4 MPa-10 psi to 200 psi0.01 or 0.00310.008 or 0.0024
PM600-A2M70.07 MPa to 2 MPa10 psi to 300 psi-0.03 MPa to 2 MPa-5 psi to 300 psi0.01 or 0.00310.008 or 0.0024
PM600-A3.5M70.07 MPa to 3.5 MPa10 psi to 500 psi-0.03 MPa to 3.5 MPa-5 psi to 500 psi0.01 or 0.00310.008 or 0.0024
PM600-A7M7ATM2 to 7 MPaATM 2 to 1000 psi0 MPa to 7 MPa0 psi to 1000 psi0.01 or 0.003 10.008 or 0.0024
PM600-A10M7ATM2 to 10 MPaATM 2 to 1500 psi0 MPa to 10 MPa0 psi to 1500 psi0.01 or 0.003 10.008 or 0.0024
PM600-A14M7ATM2 to 14 MPaATM 2 to 2000 psi0 MPa to 14 MPa0 psi to 2000 psi0.01 or 0.003 10.008 or 0.0024
PM600-A20M7ATM2 to 20 MPaATM 2 to 3000 psi0 MPa to 20 MPa0 psi to 3000 psi0.01 or 0.003 10.008 or 0.0024
PM600-A28M8ATM3 to 28 MPaATM 3 to 4000 psi0 MPa to 28 MPa0 psi to 4000 psi0.01 or 0.003 20.008 or 0.0024
PM600-A35M8ATM3 to 35 MPaATM 2 to 5000 psi0 MPa to 35 MPa0 psi to 5000 psi0.01 or 0.003 20.008 or 0.0024
PM600-A40M8ATM3 to 40 MPaATM 3 to 6000 psi0 MPa to 40 MPa0 psi to 6000 psi0.01 or 0.003 20.008 or 0.0024
PM630-A70M9ATM3 to 70 MPaATM 3 to 10000 psi0 MPa to 70 MPa0 psi to 10000 psi0.01 or 0.003 20.008 or 0.0024
PM630-A100M9ATM3 to 104 MPaATM 3 to 15000 psi0 MPa to 104 MPa0 psi to 15000 psi0.012 or 0.004 20.01 or 0.003

Notes

  1. For PM600s absolute mode modules used in absolute mode, root sum square (RSS) with 0.007 % of FS (reduced to k=1 by square root of 3). 

PM600 Pressure Measurement Modules -- REVISED

  1. PM600 and PM630 modules with full scales of 28 MPa and higher use an internal barometer in the PMM to correct for changes in barometric pressure when they are used in gauge mode and as a zeroing reference when used in absolute mode, hence there is no need to RSS 0.007 % FS. ATM is any atmospheric pressure from 70 kPa to 110 kPa (10 psi to 16 psi).
  2. For absolute ranges used in gauge mode there is an additional uncertainty of ±7 Pa for dynamic barometric compensation. When combined with other uncertainties this changes the threshold uncertainty for the PM600-A100K to 0.008 % Span and for the PM600-A200K to 0.004 % Span.
  3. These modules can be used as a barometric reference module on all chassis.
  4. Compatible with 6270A
  5. Compatible with 6270A and 8270A
  6. Compatible with 6270A, 8270A and 8370A
  7. Compatible with 8270A and 8370A
  8. Compatible with 8370A

Modelos: PM600 Pressure Measurement Modules

PM600 Pressure Measurement Modules -- REVISED
For 6270A, 8270A and 8370A Pressure Controller/Calibrators

Comprar
Converse com oAssistente da Fluke
Limpar conversa
[object Object]