Product overview: 5500A/COIL | 50-Turn Current Coil
Supports the calibration of clamp-type current meters<
The 50 Turn 5500A/COIL is used as a tool for calibrating clamp type current meters that operate by two different principles -- as current transformers (AC only), and by the Hall Effect (Both AC and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. However, by using the 50 turns of the 5500A/COIL in conjunction with a current calibrator you can effectively multiply the current of the current calibrator by a factor of 50 to support the calibration and verification of these clamp-on current meters.
Specifications: 5500A/COIL | 50-Turn Current Coil
General Specifications | ||||||||||
Number of Turns | 50 | |||||||||
Maximum Current | 11A rms, continuous 20A rms, 2 minutes | |||||||||
Maximum Duty Cycle Derating | < 11A, continuous > 11A, 2 minutes ON, 8 minutes OFF | |||||||||
Maximum Voltage | 3V rms | |||||||||
Frequency of Operation |
| |||||||||
Uncertainty Due to Clamp Meter/Coil Interaction |
| |||||||||
Calibration Documentation | Statement of calibration practices included; With 17025 accredited report | |||||||||
Minimum Inside Diameter of Clamp Jaws | 2.54 cm (1 in.) |
Models: 5500A/COIL | 50-Turn Current Coil
5500A/COIL
50-Turn Current Coil