5500A/COIL | 50-Turn Current Coil

Model#:5500A/COIL
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 21.65ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • No image available

Product overview: 5500A/COIL | 50-Turn Current Coil

Supports the calibration of clamp-type current meters<

 

The 50 Turn 5500A/COIL is used as a tool for calibrating clamp type current meters that operate by two different principles -- as current transformers (AC only), and by the Hall Effect (Both AC and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. However, by using the 50 turns of the 5500A/COIL in conjunction with a current calibrator you can effectively multiply the current of the current calibrator by a factor of 50 to support the calibration and verification of these clamp-on current meters.

Specifications: 5500A/COIL | 50-Turn Current Coil

General Specifications
Number of Turns50
Maximum Current11A rms, continuous
20A rms, 2 minutes
Maximum Duty Cycle Derating< 11A, continuous
> 11A, 2 minutes ON, 8 minutes OFF
Maximum Voltage3V rms
Frequency of Operation
DC, 45-440 Hz (should not exceed rms Voltage rating)
For the 5500A Current Output: DC, 45-65 Hz, 0-11A
 65-440 Hz, 0-2.19999A
For the 5520A Current Output: DC, 45-65 Hz, 0-20A
 65-440 Hz, 0-2.99999A
 This specification assumes that a Fluke 80i-600 Clamp Meter or equivalent is attached. Other clamp meters may limit the 5500A/5520A Current Output drive capability to less than the above.
Uncertainty Due to Clamp Meter/Coil Interaction
 ± (0.25% of effective output + 0.5A), for toroidal-wound current clamps, such as the Fluke 80I and 80I-1000.
 ± (0.50% of effective output + 0.5A), for current clamps like the Fluke 80i-kw, 80i-400, 80i-410, 80i-500, 80i-1010, Fluke 31, Fluke 33, or equivalent.
Calibration DocumentationStatement of calibration practices included; With 17025 accredited report
Minimum Inside Diameter of Clamp Jaws2.54 cm (1 in.)

Models: 5500A/COIL | 50-Turn Current Coil

5500A/COIL
50-Turn Current Coil

  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 22.92ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 22.53ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.
Chat with ourFluke assistant
Clear Chat