9007 Portable Lab Dry-Well
Product overview: 9007 Portable Lab Dry-Well
Priority Gold Instrument CarePlan
Make sure to ask your account manager about the Fluke Calibration Priority Gold CarePlan
A Fluke Calibration Priority Gold CarePlan puts you in control of your downtime and in control of your business, with calibration services and other offerings.
The 9007 covers –40 °C to 140 °C. No external cooling is needed, so you get –40 °C in normal ambient. Set-point accuracy is ± 0.15 °C and stability is better than ± 0.02 °C.
This dry-well comes with a unique calibration at the full, six-inch depth of the well and at a three-inch depth for short probes. Coefficients for each calibration are stored in the dry-well and can be easily selected from the top-panel control buttons to match the length of the probe being tested.
Encased in an all-aluminum enclosure that is durable, waterproof, and meets the standards of MIL-T-28800, the 9007 comes with both RS-232 and IEEE-488 interface connections. A wide variety of inserts are available covering probe diameters from 1/16 inch (1.6 mm) to 5/8 inch (15.9 mm).
Specifications: 9007 Portable Lab Dry-Well
Specifications | |
Range | –40 °C to 140 °C at 25 °C |
Accuracy | ± 0.15 °C |
Stability | ± 0.02 °C |
Heating Times | 25 °C to 140 °C: 20 min. |
Cooling Times | 25 °C to –40 °C: 25 min. |
Stabilization | 10 minutes |
Test Wells | 19 mm dia. x 152 mm deep (3/4 x 6 in) |
Communications | RS-232 and IEEE-488 |
Enclosure | Meets Type II, Class 3, Style D requirements of MIL-T-28800 |
Power | 115 V ac (± 10 %), 5 A or 230 V ac (± 10 %), 2.4 A, switchable, 50/60 Hz, 560 W |
Size (HxWxD) | 35.1 x 27.4 x 42.9 cm (13.8 x 10.8 x 16.9 in) |
Weight | 16.3 kg (36 lb) |
NIST-Traceable Calibration | Data at –40 °C, 0 °C, 25 °C, 75 °C, and 140 °C |