Calibration Bath Selection Guide

A calibration bath is a uniform temperature enclosure with a constant temperature setting that can be adjusted manually or with automation. These calibration baths use a mixed fluid to provide excellent thermal contact for temperature calibration. They offer high stability, a large working volume, and flexibility for calibrating a variety of temperature sensors. A world-class temperature controller and 30 plus years of experience make these the choice of National Metrology Institutes and calibration laboratories around the world.

Compact calibration baths: low and high temperature 

Model Range Stability Depth Features 
633035 °C to 300 °C ±0.005 °C at 100 °C
±0.015 °C at 300 °C 
234 mm
9.25 in 
Small benchtop footprint. 
Optional cart includes storage space. 
High temperature bath
7320–20 °C to 150 °C ±0.005 °C at –20 °C 
±0.005 °C at 25 °C 
234 mm
9.25 in 
Small 9.2-liter (2.4-gallon) tank. 
Uniformity ±0.005 °C. 
7340–40 °C to 150 °C ±0.005 °C at –40 °C 
±0.005 °C at 25 °C
234 mm
9.25 in 
Low temperature calibrations. 
Metrology-level performance. 
7380–80 °C to 100 °C ±0.006 °C at –80 °C
±0.010 °C at 0 °C 
178 mm 
7 in 
Achieves –80 °C in less than 130 minutes. 
Quiet operation. 
7312–5 °C to 110 °C ±0.001 °C at 0 °C 496 mm
19.5 in 
Maintains two TPW cells. 
Compact, quiet. 
633140 °C to 300 °C ±0.007 °C at 100 °C
±0.015 °C at 300 °C 
457 mm 
18 in 
18 in of depth with just 16 liters of fluid. 
RS-232 included. 
7321–20 °C to 150 °C ±0.005 °C at –20 °C
±0.005 °C at 25 °C 
457 mm 
18 in 
Perfect for LIG thermometers with optional kit. 
Quiet operation. 
7341–45 °C to 150 °C ±0.005 °C at –40 °C 
±0.005 °C at 25 °C 
457 mm 
18 in 
Fast temperature changes. 
Access opening accommodates many thermometers. 
7381–80 °C to 110 °C ±0.006 °C at –80 °C
±0.005 °C at 0 °C 
457 mm 
18 in 
Stability of ±0.006 °C or better over full range. 
Compatible with MET/TEMP II software. 
Low temperature bath

Standard size calibration baths: Low and high temperature 

Model Range Stability Depth Features 
7060–60 °C to 110 °C ±0.0025 °C at –60 °C
±0.0015 °C at 25 °C 
305 mm 
12 in 
Reaches –60 °C with standard refrigeration. 
7080–80 °C to 110 °C ±0.0025 °C at –80 °C 
±0.0015 °C at 25 °C 
305 mm 
12 in 
Best combination of stability and ultralow temperature bath. 
7008–5 °C to 110 °C ±0.0007 °C at 25 °C 331 mm 
13 in 
Large tank for larger mass immersion. 
Maintains standard resistors. 
7011–10 °C to 110 °C ±0.0008 °C at 0 °C
±0.0008 °C at 25 °C 
305 mm 
12 in 
Self-contained refrigeration. 
Best-priced ultrastable, cooled bath. 
7012–10 °C to 110 °C ±0.0008 °C at 0 °C 
±0.0008 °C at 25 °C 
457 mm 
18 in 
Maintains up to 4 WTP cells for weeks. 
Large access: 162 x 292 mm (6.3 in x 11.5 in). 
7037–40 °C to 110 °C ±0.002 °C at –40 °C
±0.0015 °C at 25 °C 
457 mm 
18 in 
Lowest-temperature deep-well bath. 
Mercury cell maintenance bath. 
7040–40 °C to 110 °C ±0.002 °C at –40 °C
±0.0015 °C at 25 °C 
305 mm 
12 in 
Self-contained single-stage refrigeration. 
Digital controller. 
602040 °C to 300 °C ±0.001 °C at 40 °C
±0.005 °C at 300 °C 
305 mm 
12 in 
Broad range to 300 °C. 
Optional RS-232 and IEEE-488 interface. 
602240 °C to 300 °C ±0.001 °C at 40 °C 
±0.005 °C at 300 °C 
464 mm 
18.25 in 
Deep tank for SPRT or LIG thermometers. 
Optional fluid level adapter. 
High tempertures
602440 °C to 300 °C ±0.001 °C at 40 °C 
±0.005 °C at 300 °C 
337 mm 
13.25 in 
Larger access opening and tank size for higher throughput. 
High temperature bath
6050H180 °C to 550 °C ±0.002 °C at 200 °C
±0.007 °C at 500 °C 
305 mm 
12 in 
Better stability than sand baths. 
High temperature bath, low gradients. 

Special application

Model Range Stability Depth Features 
6054 50 °C to 300 °C ±0.003 °C at 100 °C
±0.005 °C at 300 °C 
610 mm 
24 in 
Maintains constant fluid level. 
6055 200 °C to 550 °C ±0.003 °C at 200 °C
±0.01 °C at 550 °C 
432 mm
17 in 
Includes LIG sighting channel. 
Very high temperature bath
7007 –5 °C to 110 °C ±0.001 °C at 0 °C
±0.003 °C at 100 °C 
610 mm
24 in 
Large, 7-inch-diameter working space. 
7009 0 °C to 50 °C ±0.0007 °C at 25 °C 331 mm
13 in 
Largest capacity with 4.8-cubic foot (167-liter) working area and 0.7 mK stability. 
7015 0 °C to 50 °C ±0.0007 °C at 25 °C 331 mm
13 in 
Ultrastable for maintaining resistors. 
Large access and workspace. 
Splash- and spill-resistant lid. 
7108 20 °C to 30 °C ±0.004 °C 203 mm
8 in 
Peltier cooling means no compressor and quieter performance. 
Maintains standard resistors. 
7911A20 °C ±0.002 °C 203 mm
8 in 
Easy and affordable zero-point source for calibrating temperature sensors. 

Other 

Item Description 
Bath AccessoriesStands, rods, and clamps to suspend and support your probes and thermometers
Bath Fluids Silicone oils, salt, and cold fluids in convenient, small quantities. 
Rosemount Bath Controllers Model 7900 controller designed by Hart integrates the features of Hart’s 2100 controller and can be used in place of the Rosemount 915 controller with Rosemount-designed baths. 
Hart Bath Controllers Model 2100 and 2200 controllers can be integrated with homemade baths or other heat sources to achieve performance levels approaching Hart baths. 
Note: Look here for portable Micro-Baths. 
Converse com oAssistente da Fluke
Limpar conversa
[object Object]