Fluke 714B Thermocouple Calibrator

  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 115.58ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • Fluke 714B Thermocouple Calibrator
  • Fluke 714B Thermocouple Calibrator
  • Fluke 714B Thermocouple Calibrator
  • Fluke 714B Thermocouple Calibrator

Key Features

  • Highly accurate, easy-to-use thermocouple temperature calibrator
  • Measures and simulates 17 thermocouple types and millivolts
  • Measures 4 to 20 mA signals while sourcing a temperature signal
  • Comes with a magnetic hanger for convenient hands-free operation
  • Features configurable source settings for quick linearity checks

Product overview: Fluke 714B Thermocouple Calibrator

High accuracy thermocouple temperature calibration the easy way

Fluke 714B is a handheld, battery-operated instrument that measures and sources a variety of thermocouple types and millivolts. It can calibrate a linear thermocouple transmitter with the mV source function and also measure mA while sourcing temperature. It also provides linear ramp and 25% step auto ramp functionality based on 0% and 100% settings for performing quick, linearity checks.

Designed for calibration professionals

With the intelligent shutdown feature the Fluke 714B remembers settings when powered down so you can quickly replicate tests after powering back up. The calibrator also has one-year and two-year specifications, along with a traceable certificate of calibration. Combine that with its rugged construction, a backlit display, and easy to use interface and you have an ideal test tool for temperature calibration professionals who want a highly accurate, single–function thermocouple temperature calibrator.

Related resources

What's in the box:

  • 714B TC calibrator
  • Protective Boot (installed)
  • (4) AA batteries (installed)
  • Traceable calibration certificate w/data
  • (1 pair) TL75 Test Leads
  • (1 set) AC175 Alligator clips
  • Type K beaded TC
  • TPAK80 magnetic hanger strap
  • 3-year warranty
  • Quick Reference Guide

Specifications: Fluke 714B Thermocouple Calibrator

General Specifications
Maximum voltage applied between any terminal and earth ground or between any two terminals30 V
Operating temperature-10°C to 50°C
Storage temperature-30°C to 60°C
Operating altitude2,000 m
Storage altitude12,000 m
Relative humidity (% RH operating without condensation)Non condensing
90% (10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
(Without condensation)
Vibration requirementsMIL-T-28800E, Class 2
Drop text requirements1 m
IP RatingIEC 60529: IP52
Electromagnetic environmentIEC 61326-1, Portable
SafetyIEC 61010-1, Max 30 V to earth, Pollution Degree 2
Power supply4 AA NEDA 1.5A IEC LR6 batteries
Size (H x W x L)52.5 x 84 x 188.5 mm
Weight515 g
DC mA Measurement
ResolutionRangeAccuracy (% of reading + counts)
0-24 mA0.001 mA0.010% + 2 μA
Temperature coefficient±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C)
Millivolt Measurement and Source
ResolutionRangeAccuracy (% of reading +counts)
-10 mV to 75 mV0.01 mV0.015% + 10 μA
Temperature coefficient±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C)
Thermocouple Input and Output
TC TypeRange (°C)Measure (°C)Source (°C)
1 year2 year1 year2 year
E-250 to 200°C1.32.00.60.9
-200 to -100°C0.50.80.30.4
-100 to 600°C0.30.40.30.4
600 to 1000°C0.40.60.20.3
N-200 to -100°C1.01.50.60.9
-100 to 900°C0.50.80.50.8
900 to 1300°C0.60.90.30.4
J-210 to -100°C0.60.90.30.4
-100 to 800°C0.30.40.20.3
800 to 1200°C0.50.80.30.3
K-200 to -100°C0.71.00.40.6
-100 to 400°C0.30.40.30.4
400 to 1200°C0.50.80.30.4
1200 to 1372°C0.71.00.30.4
T-250 to -200°C1.72.50.91.4
-200 to 0°C0.60.90.40.6
0 to 400°C0.30.40.30.4
B600 to 800°C1.32.01.01.5
800 to 1000°C1.01.50.81.2
1000 to 1820°C0.91.30.81.2
R-20 to 0°C2.32.81.21.8
0 to 100°C1.52.21.11.7
100 to 1767°C1.01.50.91.4
S-20 to 0°C2.32.81.21.8
0 to 200°C1.52.11.11.7
200 to 1400°C0.91.40.91.4
1400 to 1767°C1.11.71.01.5
C0 to 800°C0.60.90.60.9
800 to 1200°C0.81.20.71.0
1200 to 1800°C1.11.60.91.4
1800 to 2316°C2.03.01.32.0
L-200 to -100°C0.60.90.30.4
-100 to 800°C0.30.40.20.3
800 to 900°C0.50.80.20.3
U-200 to 0°C0.60.90.40.6
0 to 600°C0.30.40.30.4
BP0 to 1000°C1.01.50.40.6
1000 to 2000°C1.62.40.60.9
2000 to 2500°C2.03.00.81.2
XK-200 to 300°C0.20.30.20.5
300 to 800°C0.40.60.30.6
G100 to 300°C1.62.41.21.8
300 to 1500°C1.01.51.01.5
1500 to 2320°C2.03.01.62.4
D0 to 300°C1.62.41.21.8
300 to 1500°C1.01.51.01.5
1500 to 2315°C2.03.01.62.4
P0 to 1000°C1.62.40.60.9
1000 to 1395°C2.03.00.81.2
M-50 to 100°C1.01.50.40.6
100 to 1000°C1.62.40.60.9
1000 to 1410°C2.03.00.81.2

Models: Fluke 714B Thermocouple Calibrator

Fluke 714B
Fluke 714B Thermocouple Calibrator

Buy it

Includes:

  • Magnetic hanging tool
  • Batteries
  • Manual
  • Traceable calibration certificate
  • Test Leads

I am trying to use the 714B Thermocouple Calibrator as a K type SOURCE. Why does the calibrator have 30mv peak to peak spikes at 1.42KHz on it's output? Spikes are a large part of the signal. Should be DC. Same issue, 2 units.

Asked by: Old_Timer
Please contact Fluke's tech support team at tech.support@fluke.com or 800-443-5853.
Answered by: FlukeCorporation
Date published: 2023-10-19
  • y_2024, m_11, d_19, h_8CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_1
  • loc_en_US, sid_3840954007604, prod, sort_[SortEntry(order=LAST_APPROVED_ANSWER_SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_fluke
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 122.59ms
  • QUESTIONS, PRODUCT
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 23.07ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.
Chat with ourFluke assistant
Clear Chat