Fluke 3130-G2M Portable Pressure Calibrator
Key Features
- Measure/generate pressures from vacuum to 2 MPa (300 psi, 20 bar).
- Measure/generate 4 to 20 mA, measures 0 to 30 V DC
- Generate vacuum to -80 kPa (-12 psi, -0.8 bar) or pressure to 2 MPa (300 psi, 20 bar).
- External gas supply up to 2 MPa (300 psi, 20 bar).
- Fine pressure adjustment.
- Measurement accuracy of ± (0.025 % reading + 0.01 % FS).
Product overview: Fluke 3130-G2M Portable Pressure Calibrator
Pressure generation and control
The internal, motorized pump provides an easy, efficient alternative to time consuming, tiring hand pumps. For applications that involve filling a large volume with pressure, the 3130-G2M allows for connection to an external gas supply such as compressed plant air. The pressure can be fine-tuned using the variable volume.
Pressure measurement
The 3130-G2M offers an onboard pressure sensor with a full scale of 2 MPa (300 psi, 20 bar) and an accuracy of ±0.025% reading + 0.01% FS (includes precision, one-year stability, uncertainty of the calibration standard, and temperature effects). In addition, the 3130-G2M can be used with the Fluke 700P pressure module series to improve measurement performance across the entire range.
Electrical measurement
The 3130-G2M offers electrical measurement capabilities for calibrating pressure transducers and transmitters, including measurements of 4 to 20 mA or 0 to 30 V dc. In addition, the 3130-G2M provides 24 V dc supply to power the device under test and can generate current in the 4 to 20 mA range.
Portability
The 3130-G2M is built into a ruggedized case with internal storage for power supply, test leads, and fittings. In addition, it includes a rechargeable, NiMH battery, allowing for approximately 50 hours of operation. When using the internal pump, the battery is sufficient to provide up to 100 calibration cycles to 300 psi. The battery is recharged by simply plugging the unit in using the included universal mains adapter.
Automation
In the field, the 3130-G2M can be connected to a personal computer with the Fluke Calibration 700SC Serial Interface Cable Assembly to log data and send instructions using Fluke Calibration COMPASS for Pressure Software. To learn more about the cable assembly and software, contact your local Fluke Calibration sales representative.
Specifications: Fluke 3130-G2M Portable Pressure Calibrator
Environmental | |
Operating temperature | -10°C to +50°C |
Storage Temperature | -20°C to +60°C |
Power requirements | 12 V DC (Universal ac adapter/charger supplied) |
Battery | Internal 3800 mAh advanced NiMH pack |
Operating Time | |
On full charge | Approx 50 hours (Measure only or external air; no pump) |
Using internal pump | Approx 100 calibration cycles to 300 psi |
Physical | |
Dimensions (L x W x D) | 15.25 x 12 x 7 in |
Weight | ~7 kg (15 lb) |
EMI/RFI conformance | EN61326:2006 Annex A |
Connectors/ports | 1/8 in NPT (External supply port and test port) |
Included accessories | Manual, NIST-traceable certificate, test leads, universal ac adapter/charger |
Ranges | |
Pressure (internal pump) | -80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar) |
Pressure (external air) | 0 to 2 MPa (0 to 300 psi, 0 to 20 bar) |
mA | 0 to 24.000 mA |
Volts | 0 to 30.000 V DC |
Engineering units | psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4°C, mmH2O @ 20°C, cmH2O @ 4°C, cmH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, mmHg @ 0°C, inHg @ 0°C |
Instrumental Measurement Uncertainty | |
Pressure | 0.025% of reading ±0.01% FS |
mA | 0.015% of reading ±0.002 mA |
Volts | 0.015% of reading ±0.002 V |
Temperature Effect (All Functions) | |
No effect on accuracy on all functions from 15°C to 35°C | |
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C |