COMPASS® for Pressure Version 3 Release Notes

Download COMPASS® for Pressure

Version 3 Releases
3.00.43 (SP1) (2011.02.11)
ADDCP-802Support PG7000 and PG9000 models that respond with FLUKE as the manufacturer.
CHGCP-793An automated PG7000/PG9000 system with a PPC and AMH will use ATM as the minimum pressure when exercising in a test that is run in Absolute by ATM.
3.00.42 (SP1) (2010.07.07)
ADDCP-689Support alternate units of measure and temperature references for the Piston Cylinder effective area.
ADDCP-775Added new instruments to the default database.
CHGCP-786When [Tools][Feature Options Display]<\ Support Temperature\> is disabled, use the DUT temperature device unit of measure to log temperature data. When the feature is enabled, the temperature unit selected during test initialization defines the unit of measure for the DUT temperature device.
CHGCP-785The Final Output Label of a device is displayed on the manual entry dialog to assist in the entering of manual data.
CHGCP-781ATM is a supported target for the pressure exercise min or max targets. Click the label next to the field to insert an ATM point.
CHGCP-766The build information in addition to the Major and Minor revisions is displayed on the Splash screen during program initialization.
CHGCP-753Font size is limited on device run screens to prevent fonts that are too large or small.
CHGCP-708[Help][About] dialog uses the new "Fluke- DH Instruments" logo.
CHGCP-691A device is now considered "in tolerance" if the error is equal to the tolerance. Previously, a device was "out of tolerance" when the error was equal to the tolerance.
CHGCP-159When a macro changes the SN or ID of a device it should also be changed on the various device displays.
CHGCP-155Additional COMPASS specific fields are exported to MET/TRACK. See Help file for complete set of fields that are mapped.
CHGCP-657The number of test cycle points can be changed from a macro.
CHGCP-751When the Data File Name selection in [Tools][Options][Data File] is set to secondary SN, the device serial number is used when the selection does not apply.
FIXCP-784Might encounter "Syntax error near '0'" when exporting test data file to MET/TRACK.
FIXCP-783Status field logged in the data file does not always update with status codes.
FIXCP-779The measurement mode of a profile DUT is not recalled between tests.
FIXCP-778Incorrect reference temperature values logged during automated test with an FPG reference.
FIXCP-762Export to MET/TRACK, data file resolution was not maintained.
FIXCP-761Export to MET/TRACK, option
FIXCP-760Removed redundant error messages associated with data files that cannot be exported to MET/TRACK.
FIXCP-759A macro run at the end of a test cycle can result in duplicate points made available to a test macro. Data file information is now synchronized prior to calling any test event macros at the end of a test cycle to avoid this problem.
FIXCP-756The COMPASS Ready option in a Test Definition would fail to return Ready if the reference and controller used different units of measure.
FIXCP-755Output resolution of DHI devices does not always conform to the front panel when the "...Preserve front panel resolution.." option is selected.
FIXCP-754Interface macros had a maximum limit of 4 calls per second regardless of the device defined poll interval value.
FIXCP-752The label associated with values that are manually entered during a tests were not always correct. In particular, when the raw output type and the final output type were different. For example, Ohms and Temperature.
FIXCP-747Target pressure on a PPC could have excessive resolution. The target is now limited to the resolution of the device.
FIXCP-744Exporting a COMPASS data file to the Test Data Database might fail if no device information is present in the data record.
FIXCP-735Line pressure and temperature fields display on final test initialization step even when the options are disabled.v FIX,CP-719 An error message would display when creating a user defined device when the Output Relationship was set to "Same". The error indicated that a unit must be selected from a drop down list when in fact the list was empty.
FIXCP-709Data viewer now shows only relevant columns on the Data grid, with the appropriate column headings.
FIXCP-703Effective area unit conversion was performed with limited resolution. This impacted the final area determination which resulted in a faulty piston gauge pressure output.
FIXCP-695BSL linearity calculation was incorrect.
FIXCP-693If a test point is removed by a macro and the test definition has multiple cycles of line pressure or temperature, the test would loop indefinitely.
FIXCP-157Shared command macro information was not exported to external databases using the Database Maintenance Tool.
FIXCP-659The Piston Gauge Platform editor displays "PC/MS" tab 2 times. The second tab should display "Sources" instead.
FIXCP-653Newly created Mass Bells are not listed in the Piston gauge Editor Mass Bell list.
FIXCP-650The True Mass label on the Mass Set Editor is missing. The field displays "kg" only.
FIXCP-644Controller list for line pressure includes 2 "Manual Entry" choices.
Chatten Sie mit unseremFluke Assistenten
Chat löschen