RPM4 v1.01I Embedded Software

RPM4 v1.01I Embedded Software (.bin)

TITLEVERUSE WITHNOT FOR USE WITH
RPM4™1.01I
2019.04.28
RPM4Note: If you have a special RPM4
model use these links:

RPM4-AD™
RPM4-AF™
RPM4-E-DWT A70M/A7M-AF™
RPM4-E-DWT™
Please contact our Calibration and Repair department at 602.773.4782 if you recently upgraded to version 1.01f or plan to upgrade from 1.01f to another version. The upgrade path from 1.01f may cause a potentially significant metrological error.
Instructions:
Download the file.

Run Winflash and select the file downloaded as the file to write to the product.

Make sure the correct COM port of the product is connected to a RS232 port on the host PC. Use a null modem cable to COM2 port of PG7000 Piston Gauge. Use a standard RS232 cable to COM1 port of PPC, PPC, RPM and molbox.

If your product has an model number ending in "-AF", it requires specific "-AF" embedded software.

The updating of embedded software should only be performed after given specific instructions to do so from a Fluke Calibration Authorized Service Center ».

If the update isn't performed properly, your product will be inoperable until it is.
Updates Include:
Ver 1.01l (2019 Apr 24)
Fixed issues related to remote commands HEAD=, READRATE and READYCK.
Fixed issues related to using the RPM4 as an external device with PPC controllers.

Ver 1.01k (2015 Aug 13)
FIX: (RPM-30) Factory remote command changes.

Ver 1.01j (2010 Sept 10)
ADD: (RPM-7): RPM4 is now compatible with alternate front panel display hardware.
FIX: (RPM4-3) Differences in US and SI version ranges prevented the use of the full range of the RPM4.
FIX: (RPM4-2) A BA100K RPT could not be auto ranged using remote commands. The issue is now resolved.

Ver 1.01i (2010 Aug 05)
CHG: (RPM-5) Internal release to support new front panel keypad.

Ver 1.01h (2009 Feb 09)
FIX: When powered on in HL (parallel) mode, the advanced Q-RPT characterization is not applied
for both IH and IL QRPTs.

Ver 1.01g (2008 Sept 08)
FIX: Full Q-RPT characterization is not being applied to pressure values.

Ver 1.01f (2008 Jul 30)
FIX: Unit change using the remote "UNIT=" command should not change the current measurement mode.
FIX: Cycling power with an unsaved AutoRange active results in odd behavior.
FIX: Absolute mode AutoZero via COM2 using an RPM4 results in an Pstd value of “0.000" being displayed instead of the reference RPM4's pressure.
FIX: RPM4 can boot to FATAL EXCEPTION BUS ERROR.
FIX: After the power up version screen, the screen goes blank for about 5 seconds before the main pressure measurement "Run screen” appears.

Ver 1.01e (2005 Nov 01)
FIX: Torr and mTorr unit conversions.

Ver 1.01d (2005 Apr 04)
Not for public release.

Ver 1.01c (2005 Mar 14)
ADD: pressure units "Torr" and "mTor"
CHG: Update the Head correction calculation to  ensures that the gauge pressure reads "0.000" after AutoZ with an Air Head applied.
FIX: Front panel menu locks when the UP or DN arrow key is pressed in a menu with selections that only span the bottom line.

Ver 1.01b (2004 Aug 25)
ADD: New G15K Q-RPT range. ADD: Rate is included in the reply to the "CONT" remote query.
FIX: Allow correct selection of pressure unit kcm2 FIX: The query character "?" is removed from the reply to query forms of the remote commands "*RST?", "ABORT?", "REMOTE?" and "LOCAL?".
FIX: RPM4 to stay in remote mode during and after use by a PPC3 until the power is cycled.
FIX: "CALAMB" command was missing the cal date in the reply.
FIX: Internal real time clock could not properly set the date past year 2009.
FIX: "GPIB" command now accepts correct address values from 1 to 30. Also, It now replies the correct address when changing the address.
FIX: Third field of the reply from the remote command "RPT" is now the RPT serial number instead of the RPM4's serial number.

Ver 1.01 (2003 May 12)
ADD: Support for use by PPC3 as an external pressure reference
ADD: Parallel measurement mode, "HL" Q-RPT functionality.
ADD: Differential mode
ADD: L2 and L3 commands
FIX: "COM1=" command could cause data corruption and a reset "ALL" on the next power cycle.
FIX: "COM1=" command reply may not be sent complete when setting a new baud rate
FIX: "LL=" command reply
FIX: Ensure that Auto-ranged UL values do not exceed the RPT's default range UL values
Chatee con nuestroasistente de Fluke
Borrar el chat