96270A RF Reference Source Firmware Upgrade

96270A RF Reference Source Firmware Upgrade (.ldr)

Serial Number RangeBackward CompatibilityDateFirmware RevisionExplanations
AllAll previous versions 96270AApril 2, 2019V1.13

Extend capability of previous 96270A models to accept Rohde & Schwarz NRPxxT power sensors.

Full list of all sensors recognised by 96270 with install firmware V1.13 installed:

Older NRP-Z51, NRP-Z52, NRP-Z55, NRP-Z56, NRP-Z57and NRP-Z58

Now extended to include:

New: NRP18T, NRP18TN, NRP33T, NRP33TN, NRP40T, NRP40TN, NRP50T, NRP50TN, NRP67T, NRP67TN and NRP110T.

Resolves minor resolved issues

Firmware Installation Instructions

  1. Power off the Product.
  2. On the back panel locate the dip switches and set dip switch 6 to its up position.
  3. Insert the USB stick.
  4. Turn on the Product. A yellow screen with an empty progress bar should appear on the screen.
  5. Follow the on screen instructions.
  6. When complete, place the dip switch (6) down again to its original position, remove the USB stick and power cycle.
  7. Firmware update complete.
Chat with ourFluke assistant
Clear Chat