7050/7050i Digital Pressure Indicators
Key features
- Ranges from 0 to 10 inH2O and 0 to 1,500 psi (0 to 25 mbar and 0 to 100 bar).
- Models: 7050i, 7050, and 7050LP.
- 7050i provides advanced precision of 0.005% of reading
- 7050 provides 0.003% of full-scale precision.
- Stability: 0.0075% of reading per year.
Product overview: 7050/7050i Digital Pressure Indicators
The Series 7050 digital pressure indicator expands upon the popular line of Series 7000 products such as the Series 7250 digital pressure controller and Series 7252 dual output pressure controller. The 7050 provides an instrument for applications that only require pressure measurement.
Three models are available to meet a wide range of applications:
- Model 7050i with advanced percent of reading precision
- Model 7050 with a high performance-to-price ratio
- Model 7050LP for low pressure measurements
The Model 7050i, 7050 and 7050LP all use a unique quartz sensor, the most accurate pressure sensing technology available. Each quartz sensor is manufactured and tested to provide the ultimate performance required by a Fluke Calibration pressure calibrator, ensuring that every customer receives quality, precision and long term stability in their instrument.
Standard precision
The Model 7050 offers an economical approach to high-accuracy pressure measurement, with a precision of 0.003 % of full scale. Various ranges from 0 to 5 to 0 to 1,500 psi (0 to 400 mbar to 0 to 100 bar) are available. For absolute mode operation, select either the barometric reference option, or the vacuum reference option, which requires an external vacuum pump connected to the reference port. The latter features an on-board vacuum sensor that allows automatic zeroing in absolute mode. Permanent absolute ranges to 50 psia (4 bar) are also available.
Advanced precision
The Model 7050i offers advanced percent of reading precision for increased capability with a single instrument, reducing the investment required to measure a wide pressure range. This model provides 0.005 % of reading precision from 25 % to 100 % of the instrument’s range. For pressures below the lower threshold of 25 %, the 7050i precision is 0.005 % of the lower threshold value. For example, a 100 psi (7 bar) Model 7050i provides 0.005 % of reading from 25 to 100 psi (1.7 to 7 bar); the precision for pressures from 0 to 25 psi (0 to 1.7 bar) is 0.005 % of 25 psi (1.7 bar). This unmatched precision is achieved through unique quartz pressure sensing technology and multiple quartz sensors in a single instrument. Various full scale pressure ranges from 0 to 5 to 0 to 1 500 psi (0 to 400 mbar to 0 to 100 bar) are available. For absolute mode operation, select either the barometric reference option, or the vacuum reference option, which requires an external vacuum pump connected to the reference port. The latter features an on-board vacuum sensor that allows automatic zeroing in absolute mode. Permanent absolute ranges to 50 psia (4 bar) are also available.
Long term stability
The Model 7050 not only provides unequalled precision, but also excellent long term stability of 0.0075 % of reading per year due to the inherent properties of quartz.
The Model 7050i not only provides unequalled precision, but also excellent long term stability of 0.0075 % of reading per year due to the inherent properties of quartz.
Automating pressure test and calibration
The7050 is provided with both an RS-232 and IEEE-488 interface, and Series 7050 syntax follows SCPI protocol for easy programming. As a standard feature, software written for previous generation Series 7215, 7010, 7000 and 6000 instruments is fully supported by the Series 7050. Intecal, an off-the-shelf software package is available. Firmware updates can be performed over the RS232 interface. A MET/CAL® driver is also available as an option.
The 7050i is provided with both an RS-232 and IEEE-488 interface, and Series 7050 syntax follows SCPI protocol for easy programming. As a standard feature, software written for previous generation Series 7215, 7010, 7000 and 6000 instruments is fully supported by the Series 7050. Intecal, an off-the-shelf software package is available. Firmware updates can be performed over the RS232 interface. A MET/CAL® driver is also available as an option.
Versatility
The Series 7050 is versatile enough to handle almost any type of pneumatic pressure measurement application.
Wide pressure range: available in a variety of standard or custom full scale pressure ranges from 10 inH2O to 1 500 psi (25 mbar to 100 bar)
Pressure units/scales: select from over twelve standard units of measure, including inHg at 0 °C and 60 °F, kPa, bar, psi, inH2O at 4 °C, 20 °C and 60 °F, kg/cm2, mmHg at 0°C, cmHg at 0 °C, and cmH2O at 4 °C, and two user defined units Head pressure: automatic correction for head pressure differences.
Absolute mode: The 7050i and 7050 offer three different methods to make absolute pressure measurements. The barometric reference option provides the most convenient method and is available on ranges 15 psi (1.0 bar) and higher. Alternatively, the vacuum reference option allows the connection of an external vacuum pump to the reference port of the instrument. An on-board vacuum sensor monitors the reference vacuum and allows for automatic zeroing in absolute mode. This option provides the lowest overall uncertainty since it does not include the additional uncertainty of a secondary barometric reference sensor. For pressures to 50 psia (4 bar), permanent absolute models are also available.
Pressure limits: Set upper and lower pressure limits to sound an audible alarm. Automatic zeroing: All models feature automatic zeroing, including units with the vacuum reference option, for automated absolute mode zeroing through the front panel or over the PC interfaces.
Options
The following options are available for the Series 7050:
- Vacuum (negative gauge) mode for bidirectional measurements
- Barometric reference for absolute mode operation with the 7050i and 7050; ranges 15 psi (1.0 bar) and higher
- Vacuum reference for absolute mode operation via an external vacuum pump connected to the reference port for 7050i and 7050
- Permanent absolute ranges to 50 psia (4 bar) full scale, which include a tare feature for simulated gauge mode operation
- NVLAP-accredited calibration report
- Intecal software for Windows® The Series 7050 digital pressure indicators provide high performance pressure measurement with a wide variety of pressure ranges and options. All are easy to use, easy to maintain, and have the reliability, performance and features that you want.
Specifications: 7050/7050i Digital Pressure Indicators
Standard Pressure Ranges | |
Model 7050 | Select any Full Ccale pressure range from 5 to 1500 psig (400 mbar to 100 barg). Or, permanent absolute ranges are available from 15 to 50 psia (1 to 4 bara). |
Optional Modes | Absolute using barometric reference sensor for ranges from 15 to 1500 psig (1.0 to 100 barg) Absolute using vacuum reference option for ranges 5 to 1500 psig (400 mbar to 100 barg) Negative gauge |
Model 7050i | Select any full scale from 5 to 1500 psig (400 mbar to 100 barg). Or, permanent absolute ranges are available from 15 to 50 psia (1 to 4 bara). |
Optional Modes | Absolute using barometric reference sensor for ranges from 15 to 1500 psig (1.0 to 100 barg) Absolute using vacuum reference option for ranges 5 to 1500 psig (400 mbar to 100 barg) Negative gauge |
Performance | |
Precision | Model 7050 0.003% of FS Model 7050i From 25% to 100% full scale (FS): 0.005% of Reading Below 25% FS: 0.005% of 25% FS Precision is defined as the combined effects of linearity, repeatability and hysteresis throughout the operating temperature range. Expression of total uncertainty conforms with the recommendations of the ISO Guide to the Expression of Uncertainty in Measurement. |
Stability | All models: 0.0075% of reading per year |
Display Resolution | User selectable to 1:1,000,000 |
Negative Gauge Precision (optional) | Model 7050 0.003% of FS Model 7050i 0.005% of 25% FS or 0.0005 psi (0.034 mbar), whichever is greater |
Barometric Reference (optional) | 0.002 psi (0.1379 mbar) maximum error per year |
Vacuum Reference (optional) | 0.0002 psi (0.014 mbar) maximum error per year |
Calibration | |
Calibration | An ISO-17025 accredited calibration is available. |
Total Uncertainty | |
Maximum deviation from the true value of pressure including precision, stability, temperature effects and the calibration standard | |
Model 7050 | Ranges to 1500 psi (100 bar) 90 day RSS: 0.003% FS + 0.002% of reading 1 year RSS: 0.003% FS + 0.0075% of reading |
Model 7050i | 25% to 100% FS 90 day: 0.006% reading 1 year: 0.009% reading |
Communications | |
Communications | RS232 and IEEE-488, SCPI syntax. Series 7250, 7215, Model 7000 and Series 6000 emulation are standard. MET/CAL® drive—optional LabView® drive Firmware updates are performed via RS232 Interface |
Languages | |
Menus and functions | The Series 7050 is capable of displaying menus and functions in: English, French, Chinese, German, Japanese, Spanish and Italian |
Options | |
Options |
*Requires external vacuum pump |
General | |
Display | TFT, VGA, active matrix, 6.4 in (162.5 mm) 640 x 480 resolution, 65,000 colors |
Temperature | Operating: 64°F to 96°F (18°C to 36°C) Storage: -4°F to 158°F (-20°C to 70°C) |
Humidity | 5% to 95% relative humidity, non-condensing |
Dimensions (h x w x d) | All versions: 7 in x 16.5 in x 19 in (178 mm x 419 mm x 483 mm) |
Weight | All models: 16 lb (7.5 kg) |
Power | 90 to 260 VAC, 50/60 Hz, 150W |
Test Port and Reference Connection | 1/4 in NPT female |
Warm Up Time | 2 to 3 hours; may be left on indefinitely |
Pressure Medium | Nitrogen or clean dry air |
Absolute using barometric reference sensor for ranges from 15 to 1500 psig (1.0 to 100 barg) Absolute using vacuum reference option for ranges 5 to 1500 psig (400 mbar to 100 barg) Negative gauge | |
25% to 100% FS 90 day: 0.006% reading 1 year: 0.009% reading |