The Fluke Calibration 9141 Dry-Well is now obsolete and has been replaced with the 9144 Field Metrology Well.
Top reasons to consider the 9144 Field Metrology Well over the 9141 Dry-Well
- Better accuracy, stability, and radial uniformity
- Dual heater zones improve vertical temperature gradient
- Greater immersion depth
- Modern user interface and industrial design
- Process Electronics option to read external reference probes and UUTs
Comparing the 9141 Dry-Well with the 9144 Field Metrology Well (with and without Process Electronics)
9141 Dry-Well | 9144 Field Metrology Well | 9144-P Field Metrology Well with Process Electronics | |
Built-in thermometer readout | No | No | Yes |
Temperature range | 50 °C to 650 °C | 50 °C to 660 °C | 50 °C to 660 °C |
Accuracy (at max) | ± 1.0 °C | ± 0.5 °C | ± 0.07 °C |
Stability (at max) | ± 0.12 °C | ± 0.05 °C | ± 0.05 °C |
Radial uniformity (at max) | ± 0.12 °C | ± 0.10 °C | ± 0.10 °C |
Axial uniformity (at max) | No spec | ± 0 5 °C | ± 0.5 °C |
Heating speed (C/min) | 52 | 41 | 41 |
Cooling speed (C/min) | 22 | 17 | 17 |
Immersion depth (mm) | 124 | 150 | 150 |
Learn more about the 9144 Field Metrology Well