Calibrate 4 sensors in half the time
Compare the 7109A Portable Calibration Bath with a 7103 Micro-Bath
A typical 3-point calibration of 4 tri-clamp sanitary sensors takes over 4 hours (259.45 minutes) using a 7103 Micro-Bath. The same calibration using a 7109A Portable Calibration Bath takes about 2 hours (130.0 minutes). That’s a time savings of more than 2 hours per 4-sensor batch!
This table presents the details | ||||
7109A Calibration Time Savings Estimate Calibrate four tri-clamp sensors at three temperature points (0, 70, and 125°C) 7103 Micro-Baths | ||||
Setpoint | 0°C | 70°C | 125°C | |
Ramp range | 23 to 0°C | 0 to 70°C | 70 to 125°C | |
Speed to temperature | 25°C to -25°C: 0.9 min/°C | 25°C to 100°C: 0.47 min/°C | 25°C to 100°C: 0.47 min/°C | |
Ramp Speed (minutes/°C) | 0.9 | 0.47 | 0.47 | |
Temperature | ||||
Start | 23°C | 0°C | 70°C | |
End | 0°C | 70°C | 125°C | |
Span | 23°C | 70°C | 55°C | |
Stabilization Time (minutes) | 60 * | 60 * | 60* | |
Calibration Time (minutes) | 80.7 | 92.9 | 85.85 | |
Total batch calibration time | 259.45 |
7109A Portable Calibration Bath | |||||
Setpoint | 0°C | 70°C | 125°C | ||
Ramp range | 23 to 0°C | 0 to 25°C | 25 to 70°C | 70 to 125°C | |
Speed to temperature | 25°C to -25°C: 1.5 min/°C | -25°C to 25°C: 0.7 min/°C | 25°C to 140°C: 0.48 min/°C | 25°C to 140°C: 0.48 min/°C | |
Ramp Speed (minutes/°C) | 1.5 | 0.7 | 0.48 | 0.48 | |
Temperature | |||||
Start | 23°C | 0°C | 25°C | 70°C | |
End | 0°C | 25°C | 70°C | 125°C | |
Span | 23°C | 25°C | 45°C | 55°C | |
Stabilization Time (minutes) | 10 | n/a | 10 | 10 | |
Calibration Time (minutes) | 44.5 | 17.5 | 31.6 | 36.4 | |
Total batch calibration time | 130.0 | ||||
The 7109A saves over 2 hours per four sensor batch | Time savings | 129.45 | |||
*The 7103 Micro-Bath calibrates one tri-clamp sensor at a time. Each sensor takes 15 minutes to stabilize for a total of 60 minutes per batch of four sensors. |
Learn more about the 6109A and 7109A Portable Calibration Baths