5790A AC Measurement Standard

  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 48.39ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • No image available
Discontinued
This product is discontinued or reclassified

Product overview: 5790A AC Measurement Standard

Versatility that keeps you productive

When you first power up the 5790A AC Measurement Standard, diagnostics verify the instrument's integrity.

The variety of input connections allows you to use the one that best suits your application. There are four sets of input terminals on the 5790A, two Type-N connectors and two sets of five-way binding posts. One Type-N and one set of binding posts are dedicated to the ac measurement and transfer modes. AC or dc voltages may be applied to either input connection over the 5790's full range, allowing you to perform automated ac/dc transfer measurements. The 5790A determines automatically whether the applied voltage is ac or dc.

The second Type-N input connection supports the optional wideband mode, and the second set of binding posts are designed for A40 Series Current Shunts.

The input connection is selected with the touch of a key on the 5790A front panel. An LED indicates which selection is active.

Whether you are using the 5790A as a voltmeter or a transfer standard, input voltage and frequency are always indicated on the measurement display. In the transfer mode, the ac/dc or ac/ac difference is always indicated on the control display in ppm, %, volts or ratio.

The 5790A is a fully autoranging instrument and selects the best voltage range for the measurement you are making. You may also select and lock in ranges manually. Robust 1200V input protection is active on all voltage ranges.

Using the trigger keys, the 5790A can switch from continuous to single measurements of the input voltage, making it easy to take sample readings at predetermined intervals.

When using the 5790A in transfer mode, the reference voltage is stored automatically, and all ac/dc or ac/ac difference measurements are made relative to it. At any time, you can view the reference by pressing the VIEW REF key. You may also store the average of two voltages as a reference to eliminate dc reversal errors, for example.

The intuitive front panel layout of the 5790A makes manual operation fast and simple. Keys and selections are logically arranged and labeled. And messages and menus are displayed clearly on the 5790A's bright, vacuum fluorescent display.

The 5790A is at home in automated systems as well. GPIB/IEEE-488 and RS-232C interfaces are included and all functions of the instruments can be controlled by a variety of host computers, including PCs. Using an available instrument driver, the 5790A can be integrated into automated systems operating under MET/CAL® Plus Calibration Software.

Designed with your support requirements in mind

The 5790A AC Measurement Standard provides a self-contained calibration procedure designed to simplify periodic performance verification. The operator is prompted on what actions to take. To minimize the equipment required, the 5790A is designed to be supported by the 792A AC/DC Transfer Standard.

Specifications: 5790A AC Measurement Standard

General Specifications
AC V MeasurementRange: 700 mV to 1000V
10 Hz to 1 MHz
Best 1Year Specification: ± 24 ppm
Wideband MeasurementRange: 700 mV to 7V
10 Hz to 30 MHz
Best 1Year Specification: ± 0.35%
Warm up Time30 minutes
Relative HumidityOperating: 45% to 50°C; 75% to 45°C; 95% to 30°C
Storage: < 95% non-condensing
AltitudeOperating: 3,050 meters (10,000 feet)
Non-Operating: 12,200 meters (40,000 feet)
Temperature PerformanceOperating: 0°C to 50°C
Calibration: 15°C to 35°C
Storage Temperature-40°C to 70°C
EMI/RFIComplies with FCC Part 15 Subpart B, Class B; VDE 0871, Class B; ESD: EIA PN-1361. Surge: ANSI C62.41-1980, Category A
Reliability MIL-T-28800D, paragraph 3.13.3
Line Power47 Hz to 63 Hz; +/-10% of selectable line voltages: 100V, 110V, 115V,
120V, 200V, 220V, 230V, 240V
SafetyComplies with UL 1244 and IEC348-1978 and IEC 1010 and CSA C22.2 No. 231
InterfacesRS-232C, IEEE-488
Maximum Power5790A: 95 VA;
with wideband: 120 VA
Calibration Documentation17025 accredited report of calibration included
SizeHeight: 17.8 cm (7 in) standard rack mount +1.5 cm (0.6 in)
Width: 43.2 cm (17 in)
Depth: 63 cm (24.8 in)
Weight5790A: 24 kg (53 lbs.); with wideband: 24.5 kg (54 lbs

Models: 5790A AC Measurement Standard

5790A
AC Measurement Standard

  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 50.37ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 23.62ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.