CalTool for Analog Free Utility

CalTool for Analog Free Utility (.exe)

TITLEVERUSE WITHDOCUMENTS
CalTool™ for Analog Free Utility1.00
TITLEVERUSE WITHDOCUMENTS
CalTool™ for Analog Free Utility1.00
2003.01.27
molbox1 (w/MFC option installed) & MFC-CBHelp File Included
Instructions:
Download and run the file. The installation program will prompt you through the process.
Description
CalTool™ for Analog is used to calibrate the analog circuits of molbox1 and MFC-CB. CalTool for Analog is PC based application that assists in the calibration of the analog circuit of supported Fluke Calibration products. Specifically molbox1s that have the optional analog option and the MFC-CB.
Summary
CalTool for Analog is only used with molbox1s that have an MFC option installed and with an MFC CB. If you have one of these products, locate the calibration cable that was shipped with the product and obtain a DMM for calibration. Install and run CalTool for Analog, then follow the instructions in CalTools help menu.

In general, follow these steps:
1. Select the [Setup] menu choice.
2. Specify the remote interface to the product (molbox1 or MFC CB) and for the DMM.
3. Enter the desired voltage and current test points. The default values are recommended.
    Press [Accept] when ready.
4. Press the [Begin Test] button at the bottom of the main display.
5. Several prompts will follow to determine the type of calibration to perform (voltage, current or both).
6. Connect the product to the DMM as requested by CalTool and follow the program prompts.
7. CalTool will acquire data from the product and the DMM and determine new calibration coefficients.
    Options are provided to write this information to the product and to generate a basic calibration report.
Updates Include:
Ver 1.00 (03 JAN 27)
Initial software release.