Fluke 2271A Industrial Pressure Calibrator
Key features
- Everything you need to calibrate and test pressure transmitters and gauges from very low pressure to 3000 psi (20 MPa), all in one box
- Modular design supports multiple accuracy classes (0.02 % FS or 0.01 % Reading) with the ability to expand to cover wider pressure ranges
- Ideal for both beginner and expanding calibration laboratories.
- CPS Feature: Optional system to prevent contamination.
- Multilingual, intuitive interface and design.
Product overview: Fluke 2271A Industrial Pressure Calibrator
Wide workload coverage for the present and the future
The 2271A features pressure ranges from -100 kPa to 20 MPa (-15 to 3000 psi), which covers the requirements of a wide range of gauges and sensors. Thanks to its modular design, the 2271A enables you to install two modules with different measurement ranges within the same chassis. You can purchase modules to match your current workload now; later, when your workload changes and grows, you can easily add ranges, swapping modules in and out according to your workload. This flexibility enables you to maintain your investment in the 2271A for years to come.
A built-in electrical measurement module (EMM) with HART capabilities enables you to perform closed loop, fully automated calibration on 4-20 mA devices such as smart transmitters, gauges and switches. Just set up the 2271A and then walk away to attend to other tasks.
The EMM supplies 24 V dc loop power for measuring mA and V dc. It has a built-in 250 Ohm resistor that can be toggled on or off, eliminating the need to have an external resistor to enable HART communications.
The 2271A accuracy specifications are provided in full and supported by a Technical Note that details its measurement uncertainty, so you know exactly what you are getting. The Technical Note is available for download on the flukecal.com website. As with all Fluke Calibration instruments, these specifications are conservative, complete and dependable.
Versatile pressure measurement modules
The 2271A uses PM200 Pressure Measurement Modules. These modules use a highly characterized silicon pressure sensor to provide an economical method of making accurate pressure measurements. The 0.02% full scale (FS) specification includes the short-term performance of the module (linearity, hysteresis and repeatability) as well as its long-term stability and the uncertainty of the calibration standard. Users can be confident in the PM200 measurement performance.
Good: PM200 Pressure Measurement Modules | Better: PM500 Pressure Measurement Modules |
---|---|
|
|
Install up to two pressure modules in the 2271A chassis at one time, mixing and matching module ranges to get the combination that best suits your needs. There is no limit to the number of modules that can be used with the system, allowing you to change pressure ranges on the fly to meet your needs. Modules snap in and out quickly and easily through the front of the 2271A; just slide each into a specially-designed track and tighten the knob until you hear it click into place. The click tells you the module is safely in place; a special anti-torque guard on the knob prevents over-tightening so you never have to wonder if you tightened it too much or not enough.
Each module uses an enhanced face-seal design that has been leak tested to pressures three times higher than the maximum working pressure. You don’t have to worry about a leak in the system affecting your ability to measure and control pressure.
Conveniently located dual test ports and reference port
Dual test ports on the top of the 2271A let you easily mount two devices under test (DUTs). You can potentially double your throughput without spending time searching for fittings and tees. The vertical test ports let you easily connect analog dial gauges without the need for additional test stands or manifolds. Two test port types are supported, the standard HC20 or the P3000 test port. Both types of test ports enable you to make hand tight connections to traditional NPT, BSP, or metric pressure fittings. The HC20 test port includes easy grip features and integral support for M20 connections, whereas the P3000 style test port provides backwards compatibility for users of Fluke Calibration P3000 deadweight testers or P5500 pressure comparators.
A reference port is also located on top of the 2271A for applications that require an atmospheric reference.
Safety features protect operators and instruments
Each measurement module, as well as the main chassis, has pressure relief valves to protect the instrument and its operators from accidental overpressure. The 2271A has been designed using Sound Engineering Practices (SEP). With the internal relief valves, user-settable pressure limits, and emergency abort button, safety is the highest priority.
Preventing contamination
If your workload includes devices that contain different substances like water, oil and gas, you could be at risk for contamination—something getting into your system that isn’t supposed to be there. Contamination can clog a calibrator’s valves, wear out its parts, and make it difficult to maintain pressure. If the contamination gets into the sensor, it can actually change the calibrator’s behavior and throw off your readings. If contamination is a concern to you, order the optional Contamination Prevention System (CPS) to help keep the calibrator’s valves clean and free from debris.
The CPS provides an unprecedented level of protection by maintaining uni-directional flow away from the controller, a gravity sump system, and a two-stage filtering system.
Automated Pressure Testing: Automate with COMPASS® software for improved consistency and throughput
Fluke Calibration COMPASS for Pressure software is designed specifically for pressure calibration. It enables you to automate the 2271A and run complete automated pressure testing calibration sequences on single or multiple devices under test. COMPASS software removes the unknowns often associated with getting automated systems online. The 2271A also features a full remote interface that enables you to use it with custom software or other data acquisition equipment. Details about the interface are provided in the 2271A User Manual.
CarePlans help you manage cost of ownership
Reduce downtime and control your cost of ownership with a CarePlan. Fluke Calibration offers one-year, three-year and five-year Priority Gold CarePlans, which feature an annual standard or accredited calibration of your 6270A calibrator with guaranteed six-day in-house turnaround1 , plus free repairs with guaranteed ten-day in-house repair (includes calibration). One-year, three-year, and five-year Silver CarePlans are available for those customers who only want extended warranty coverage.
We’re here to help
Fluke Calibration’s testing, repair and calibration services are dedicated to filling your needs quickly and at a fair cost while maintaining the unmatched level of quality that is our trademark. Our pressure calibration laboratories are accredited to ISO/IEC 17025. We maintain global calibration and repair facilities to help you keep your hardware in top working order.
Specifications: Fluke 2271A Industrial Pressure Calibrator
General specifications | |
Power requirements | 0 V ac to 240 V ac, 47 Hz to 63 Hz |
Fuse | T2A 250 V ac |
Max power consumption | 100 W |
Operating ambient temperature range | 15 °C to 35 °C |
Storage temperature | -20 °C to 70 °C |
Relative humidity | Operating: <80 % to 30 °C, <70 % to 35 °C |
Storage: <95 %, non-condensing. A power stabilization period of four days may be required after extended storage at high temperature and humidity. | |
Vibration | MIL-T-28800D CLASS 3 |
Altitude (Operation) | <2000 m |
Ingress protection | IEC 60529: IP20 |
Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |
Warmup time | 15 minutes typical |
Weight (chassis only) | 15 kg (33.06 lbs) |
Dimensions | Height: 2271A-NPT-HC20 305 mm (12 in) |
2271A-BSP-HC20 305 mm (12 in) | |
2271A-NPT-P3K 237 mm (9.33 in) | |
2271A-BSP-P3K 237 mm (9.33 in) | |
Width: 442 mm (17.40 in) | |
Depth: 446 mm (17.55 in) |
Control specifications | ||
Control Precision (Dynamic Mode) | PM200-BG2.5K | 0.005% Range Span |
All other ranges | 0.001 % Range Span | |
Control turndown | 10:1 (Typical) | |
Low control point | 1 kPa (0.15 psi) absolute | |
Control turndown is defined as the relationship between the provided supply pressure and the appropriate supply pressure for the range. For example, a unit with a 7 MPa (1000 psi) and 700 kPa range (100 psi) with a supply pressure of 7.7 MPa (1100 psi) provides control precision of 0.001 % range because 7 MPa is 10 times greater than 700 kPa. A system with ranges of 20 MPa (3000 psi) and 700 kPa (100 psi) with supply pressure of 22 MPa (3300 psi) will have 0.001 % range control precision on the 20 MPa range but only 0.003 % control precision on the 700 kPa range. Control precision of 0.001 % on the low range can be achieved by reducing the supply pressure. |
Supply pressure requirements Clean dry air or nitrogen (industrial grade, 99.5 %) | |
Maximum particulate contamination | ≤ 1.25 micrometer (50 microinches) |
Maximum moisture content | -50 °C dew point |
Maximum hydrocarbon content | 30 ppm |
Interface/ communications | |
Primary remote interfaces | Ethernet, RS232, USB |
Electrical Measurement Module (EMM) | |
Connection | Standard 4 mm jack |
Maximum 30 V dc w.r.t. chassis ground | |
Aux drivers | 4 external solenoid drivers |
24 V dc. 100 % duty cycle when turned, reducing to 40 % shortly after. |
Pressure measurement specifications | ||||
PM200 Modules | ||||
Model | Range (SI units) | Range (Imperial units) | Measurement mode | 1 year specification (%FS) |
PM200-BG2.5K | -2.5 kPa to 2.5 kPa | -10 inH₂0 to 10 inH₂0 | gauge | 0.20 % |
PM200-BG35K | -35 kPa to 35 kPa | -5 psi to 5 psi | gauge | 0.05 % |
PM200-BG40K | -40 kPa to 40 kPa | -6 psi to 6 psi | gauge | 0.05 % |
PM200-BG60K | -60 kPa to 60 kPa | -9 psi to 9 psi | gauge | 0.05 % |
PM200-A100K | 2 kPa to 100 kPa | 0.3 psi to 15 psi | absolute | 0.10 % |
PM200-BG100K | -100 kPa to 100 kPa | -15 psi to 15 psi | gauge | 0.02 % |
PM200-A200K | 2 kPa to 200 kPa | 0.3 psi to 30 psi | absolute | 0.10 % |
PM200-BG200K | -100 kPa to 200 kPa | -15 psi to 30 psi | gauge | 0.02 % |
PM200-BG250K | -100 kPa to 250 kPa | -15 psi to 36 psi | gauge | 0.02 % |
PM200-G400K | 0 kPa to 400 kPa | 0 psi to 60 psi | gauge | 0.02 % |
PM200-G700K | 0 kPa to 700 kPa | 0 psi to 100 psi | gauge | 0.02 % |
PM200-G1M | 0 MPa to 1 MPa | 0 psi to 150 psi | gauge | 0.02 % |
PM200-G1.4M | 0 MPa to 1.4 MPa | 0 psi to 200 psi | gauge | 0.02 % |
PM200-G2M | 0 MPa to 2 MPa | 0 psi to 300 psi | gauge | 0.02 % |
PM200-G2.5M | 0 MPa to 2.5 MPa | 0 psi to 360 psi | gauge | 0.02 % |
PM200-G3.5M | 0 MPa to 3.5 MPa | 0 psi to 500 psi | gauge | 0.02 % |
PM200-G4M | 0 MPa to 4 MPa | 0 psi to 580 psi | gauge | 0.02 % |
PM200-G7M | 0 MPa to 7 MPa | 0 psi to 1000 psi | gauge | 0.02 % |
PM200-G10M | 0 MPa to 10 MPa | 0 psi to 1500 psi | gauge | 0.02 % |
PM200-G14M | 0 MPa to 14 MPa | 0 psi to 2000 psi | gauge | 0.02 % |
PM200-G20M | 0 MPa to 20 MPa | 0 to 3000 psi | gauge | 0.02 % |
Models: Fluke 2271A Industrial Pressure Calibrator
Industrial Pressure Calibrator
A complete pressure calibration system includes:
- 1 chassis (2271A-NPT-XXX or 2271A-BSP-XXX)
- 1 or more measurement modules (PM200 or PM500)
Industrial Pressure Calibrator Chassis, BSP Manifold, HC20 Test Port Connections
Industrial Pressure Calibrator Chassis, BSP Manifold, P3000 Test Port Connections
Pressure Calibrator, NPT, PM200-G20M/BG250K, 20 MPA (3000 PSI)
Pressure Calibrator, NPT, PM200-BG200K/BG2.5K, 200 KPA (30 PSI)
Manuals + resources: Fluke 2271A Industrial Pressure Calibrator
- Automating the calibration of analog output pressure transmitters using the 2271A Industrial Pressure Calibrator
- Cardinal point calibration of pressure dial gauges using the Fluke 2271A Industrial Pressure Calibrator
- Guide to determining pressure measurement uncertainty for PM200, PM500, and PM600 Pressure Measurement Modules
- Simplifying the maintenance of HART pressure transmitters with the 2271A Industrial Pressure Calibrator