5500A Multi-Product Calibrator
Discontinued
This product is discontinued or reclassified
Fluke suggests:5502A Multi-Product Calibrator
Product overview: 5500A Multi-Product Calibrator
Specifications: 5500A Multi-Product Calibrator
Technical Specifications | |
Voltage DC | Range: 0 to ±1020V Best 1 Year Specification: ±50 ppm of setting |
Current DC | Range: 0 to ±11A Best 1 Year Specification: ±0.01% of setting |
Resistance | Range: 0 to 3.29.999 MΩ Best 1 Year Specification: ±0.009% of setting |
Voltage AC | Range: 1 mV to 1020V 10 Hz to 500 kHz, Sine Best 1 Year Specification: ±0.03% of setting |
Current AC | Range: 29 m A to 11A 10 Hz to 10 kHz, Sine Best 1 Year Specification: ±0.06% of setting |
Capacitance | Range: 0.33 nF to 1.1 mF Best 1 Year Specification: ±0.25% of setting |
Thermocouple Source and Measure | Range: -250°C to +2316°C (11 types) Best 1 Year Specification: ± 0.14°C of setting |
RTD Source | Range: -200°C to +630°C (8 types) Best 1 Year Specification: ±0.03°C of setting |
DC Power | Range: 109 mW to 11 kW Best 1 Year Specification: ± 0.08% of setting |
AC Power | Range: 109 mW to 11 kW 45-65 Hz, PF=1 Best 1 Year Specification: ± 0.15% of setting |
Phase | Range: 0 to ± 179.99° Best 1 Year Specification: ±0.15° |
Frequency | Range: 0.01 Hz to 2.0 MHz Best 1 Year Specification: ±25 ppm of setting |
General Specifications | |
Warm up Time | Twice the time since last warmed up, to a maximum of 30 minutes |
Settling Time | Less than 5 seconds for all functions and ranges except as noted |
Standard Interfaces | IEEE-488 (GPIB), RS-232, 5725A (5500A only) |
Temperature Performance | Operating: 0°C to 50°C Calibration: (tcal): 15°C to 35°C Storage: -20°C to 70°C |
Temperature Coefficient | Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C |
Relative Humidity | Operating: < 80% to 30°C, < 70% to 40°C, < 40% to 50°C Storage: < 95%, non-condensing |
Altitude | Operating: 3,050m (10,000 ft) maximum Non-operating: 12,200m (40,000 ft) maximum |
Safety | Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92 |
Analog Low Isolation | 20V |
EMC | Designed to comply with FCC Rules Part 15 |
Line Power | Line Voltage (selectable): 100V, 120V, 220V, 240V Line Frequency: 47 to 63 Hz Line Voltage Variation: : ± 10% about line voltage setting |
Power Consumption | 300 VA |
Calibration Documentation | 17025 accredited report of calibration included |
Dimensions | Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit Width: 43.2 cm (17 in), standard rack width Depth: 47.3 cm (18.6 in) overall |
Weight (without options) | 20 kg (44 lbs.) |
Absolute Uncertainty Definition | 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated. |
Specification Confidence Interval | > 99% |
Models: 5500A Multi-Product Calibrator
5500A
Multi-Product Calibrator