Fluke 2600A-101 Sprt Probe
Key features
- The 2600A-101 is a precision four-wire, 100 Ohm SPRT (Secondary Platinum Resistance Thermometer) probe that have been designed and tested to meet the precision required as a secondary reference standard.
Product overview: Fluke 2600A-101 Sprt Probe
Secondary Platinum Resistance Thermometer Probe
The 2600A-101 is a precision four-wire, 100 Ohm SPRT (Secondary Platinium Resistance Thermometer) probe that have been designed and tested to meet the precision required as a secondary reference standard. The 2600A-101 has stripped leads suitable for connection to any Fluke data acquisition product or other screw terminal connector. This probe comes with calibration certificates and data for -200 to 6600°C.
Probe hook up diagram:
The probe connects to channel 1(input) and 11(excitation) on the Hydra input connector.
Warning: Probe handle surfaces are conductive. Use caution near electric circuits. Voltage levels above 30V rms, 42 V peak, or 60 V dc should not be applied to the probe.
Specifications: Fluke 2600A-101 Sprt Probe
General Specifications | |||||||||||||||||||
Use Temperature | 200 °C to 660 °C | ||||||||||||||||||
Stability | Typical draft < 0.01°C after 100 times from 25 to 660 °C | ||||||||||||||||||
Calibrated Accuracy |
| ||||||||||||||||||
Performance Specifications (nominal) |
| ||||||||||||||||||
Dimensions |
|
Models: Fluke 2600A-101 Sprt Probe
2600A-101
SPRT Probe